Mechanical and dose delivery accuracy evaluation in radiosurgery using polymer gels
Author(s) -
Sandilos Panagiotis,
Tatsis Elias,
Vlachos Lampros,
Dardoufas Constantinos,
Karaiskos Pantelis,
Georgiou Evangelos,
Baras Panagiotis,
Kipouros Panagiotis,
Torrens Michael,
Angelopoulos Angelos
Publication year - 2006
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v7i4.2273
Subject(s) - radiosurgery , dosimeter , imaging phantom , dosimetry , isocenter , nuclear medicine , collimator , radiation treatment planning , materials science , dose profile , irradiation , biomedical engineering , linear particle accelerator , medical physics , medicine , radiation therapy , physics , optics , beam (structure) , radiology , nuclear physics
The polymer gel–magnetic resonance imaging (MRI) dosimetry technique was used to evaluate the mechanical and dose delivery accuracy in Leksell gamma‐knife stereotactic radiosurgery for the treatment of multiple targets. Two different polymer gel dosimeter formulations reported in the literature were prepared in‐house. A plan for the treatment of four brain metastases (targets) was generated. It involved the delivery of four 8‐mm collimator shots using different prescription isodose lines and different prescription doses for each target, keeping the maximum dose constant for all targets. A sample of each gel formulation was irradiated using a custom‐made phantom with an experimental procedure capable of testing the increased nominal mechanical accuracy of stereotactic radiosurgery. The irradiated dosimeters were evaluated using a clinical 1.5 T MR imager. Result manipulation in 3D allowed for the determination of the mechanical accuracy in the delivery of each shot through the comparison of measured versus planned shot center coordinates. Dose delivery accuracy was also evaluated by comparison of maximum dose values measured at the center of each shot as well as dose distribution measurements, with corresponding treatment‐planning calculations. Polymer gel dosimetry was found capable of verifying the complete chain of radiosurgery treatment in gamma‐knife applications involving the irradiation of multiple targets. PACS numbers: 87.53.Dq, 87.53.Ly, 87.53.Xd

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here