Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
Author(s) -
Rodríguez Miguel L.,
deAlmeida Carlos E.
Publication year - 2002
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v3i4.2552
Subject(s) - imaging phantom , monte carlo method , brachytherapy , rectum , nuclear medicine , dosimetry , radiation treatment planning , dose rate , physics , materials science , medicine , medical physics , mathematics , radiology , radiation therapy , surgery , statistics
Monte Carlo calculations of the absorbed dose at various points of a brachytherapy anthropomorphic phantom are presented. The phantom walls and internal structures are made of polymethylmethacrylate and its external shape was taken from a female Alderson phantom. A complete Fletcher‐Green type applicator with the uterine tandem was fixed at the bottom of the phantom reproducing a typical geometrical configuration as that attained in a gynecological brachytherapy treatment. The dose rate produced by an array of five 137 Cs CDC‐J type sources placed in the applicator colpostats and the uterine tandem was evaluated by Monte Carlo simulations using the code penelope at three points: point A, the rectum, and the bladder. The influence of the applicator in the dose rate was evaluated by comparing Monte Carlo simulations of the sources alone and the sources inserted in the applicator. Differences up to 56% in the dose may be observed for the two cases in the planes including the rectum and bladder. The results show a reduction of the dose of 15.6%, 14.0%, and 5.6% in the rectum, bladder, and point A respectively, when the applicator wall and shieldings are considered. PACS number(s): 87.53Jw, 87.53.Wz, 87.53.Vb, 87.66.Xa