The overshoot phenomenon in step‐and‐shoot IMRT delivery
Author(s) -
Ezzell Gary A.,
Chungbin Suzanne
Publication year - 2001
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v2i3.2607
Subject(s) - monitor unit , overshoot (microwave communication) , head and neck , physics , nuclear medicine , computer science , mathematics , medicine , surgery , telecommunications
The control loop in the Varian DMLC system (V4.8) requires ~ 65 msec to monitor and halt the irradiation of a segment, causing an “overshoot” effect: the segment ends on a fractional monitor unit larger than that planned. As a result, the actual MU delivered may differ from that planned. In general, for step‐and‐shoot treatments, the first segment receives more, the last receives less, and intermediate segments vary. The overshoot for each segment (ΔMU) is small, approximately 0.6 MU at 600 MU/min Our IMRT planning system (Corvus) produces plans often having more than 20% of the segments with less than 1 MU/segment. Such segments may be skipped if the ΔMU exceeds the segments’ planned MU. Furthermore, QA filming often requires reducing the total MU by a factor of 4–6, increasing the potential for dosimetric error. This study measured ΔMU over a range of MU/min and MU/segment. At > 5 MU/segment, the ΔMU was stable, corresponding to a delay of 62 msec. ΔMU became larger and more variable at < 1 MU/segment. The behavior was modeled in a computer program that predicted the change in delivered MU/segment and total change in delivered MU to each beamlet. Beams were analyzed for patients receiving 5 field prostate or 9 field head and neck treatments. At 400 MU/min, 28% and 16%, respectively, of the planned segments were skipped. For QA filming, up to 75% of the segments were skipped. The cumulative error averaged < 0.1 MU/beamlet, but individual beamlets had errors exceeding 200%. The effect is most significant for low dose regions. Recommendations are given for deciding when to treat or do QA studies with lower MU/min. In general, treatments are not significantly affected, but QA films taken at reduced MU may be improved if irradiated at lowered MU/min. PACS number(s): 87.53.–j, 87.90.+y