Experimental validation of a kV source model and dose computation method for CBCT imaging in an anthropomorphic phantom
Author(s) -
Poirier Yannick,
Tambasco Mauro
Publication year - 2016
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v17i4.6021
Subject(s) - imaging phantom , nuclear medicine , thermoluminescent dosimeter , cone beam computed tomography , absorbed dose , ionization chamber , dose profile , physics , dosimetry , computed tomography , medicine , dosimeter , radiology , ionization , ion , quantum mechanics
We present an experimental validation of a kilovoltage (kV) X‐ray source characterization model in an anthropomorphic phantom to estimate patient‐specific absorbed dose from kV cone‐beam computed tomography (CBCT) imaging procedures and compare these doses to nominal weighted CT‐dose index ( CTDI w ) dose estimates. We simulated the default Varian on‐board imager 1.4 (OBI) default CBCT imaging protocols (i.e., standard‐dose head, low‐dose thorax, pelvis, and pelvis spotlight) using our previously developed and easy to implement X‐ray point‐source model and source characterization approach. We used this characterized source model to compute absorbed dose in homogeneous and anthropomorphic phantoms using our previously validated in‐house kV dose computation software (kVDoseCalc). We compared these computed absorbed doses to doses derived from ionization chamber measurements acquired at several points in a homogeneous cylindrical phantom and from thermoluminescent detectors (TLDs) placed in the anthropomorphic phantom. In the homogeneous cylindrical phantom, computed values of absorbed dose relative to the center of the phantom agreed with measured values within ≤ 2 % of local dose, except in regions of high‐dose gradient where the distance to agreement (DTA) was 2 mm. The computed absorbed dose in the anthropomorphic phantom generally agreed with TLD measurements, with an average percent dose difference ranging from 2.4 % ± 6.0 % to 5.7 % ± 10.3 % , depending on the characterized CBCT imaging protocol. The low‐dose thorax and the standard dose scans showed the best and worst agreement, respectively. Our results also broadly agree with published values, which are approximately twice as high as the nominal CTDI w would suggest. The results demonstrate that our previously developed method for modeling and characterizing a kV X‐ray source could be used to accurately assess patient‐specific absorbed dose from kV CBCT procedures within reasonable accuracy, and serve as further evidence that existing CTDI w assessments underestimate absorbed dose delivered to patients. PACS number(s): 87.57.Q‐, 87.57.uq, 87.10.Rt

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here