Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI
Author(s) -
Narayanasamy Ganesh,
Cruz Wilbert,
Saenz Daniel L.,
Stathakis Sotirios,
Papanikolaou Niko,
Kirby Neil
Publication year - 2016
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v17i3.6128
Subject(s) - dosimeter , dosimetry , nuclear medicine , bolus (digestion) , lung , materials science , block (permutation group theory) , biomedical engineering , medicine , mathematics , surgery , geometry
Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the accuracy of these lung block verification measurements. Initially, the thickness of the lung block was increased to provide adequate lung sparing. Further tests revealed the increase was due to electron contamination dose emanating from the lung block. The thickness of the bolus material covering the OSLD behind the lung block was increased to offset the electron contamination. In addition, the distance from the lung block to the dosimeter was evaluated for its effect on the OSLD reading and found to be clinically insignificant over the range of variability in our clinic. The results show that the improved TBI treatment technique provides for better accuracy of measured dose in vivo and consistency of patient setup. PACS number(s): 87.53.Bn, 87.53.Kn, 87.55.N‐, 87.55.Qr

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here