
Seed coordinates of a new COMS‐like 24 mm plaque verified using the FARO Edge
Author(s) -
Cutsinger Sarah E. McCauley,
Furutani Keith M.,
Forsman Renae M.,
Corner Stephen M.
Publication year - 2015
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v16i6.5721
Subject(s) - standard deviation , enhanced data rates for gsm evolution , mathematics , brachytherapy , biomedical engineering , computer science , medicine , statistics , artificial intelligence , surgery , radiation therapy
A 24 mm COMS‐like eye plaque was developed to meet the treatment needs of our eye plaque brachytherapy practice. As part of commissioning, it was necessary to determine the new plaque's seed coordinates. The FARO Edge, a commercially available measurement arm, was chosen for this purpose. In order to validate the FARO Edge method, it was first used to measure the seed marker coordinates in the silastic molds for the standard 10, 18, and 20 mm COMS plaques, and the results were compared with the standard published Task Group 129 coordinates by a nonlinear least squares match in MATLAB version R2013a. All measured coordinates were within 0.60 mm, and root mean square deviation was 0.12, 0.23, and 0.35 mm for the 10, 18, and 20 mm molds, respectively. The FARO Edge was then used to measure the seed marker locations in the new 24 mm silastic mold. Those values were compared to the manufacturing specification coordinates and were found to demonstrate good agreement, with a maximum deviation of 0.56 mm and a root mean square deviation of 0.37 mm. The FARO Edge is deemed to be a reliable method for determining seed coordinates for COMS silastics, and the seed coordinates for the new 24 mm plaque are presented. PACS number: 87.53.Jw