Comparative evaluation of modern dosimetry techniques near low‐ and high‐density heterogeneities
Author(s) -
Alhakeem Eyad A.,
AlShaikh Sami,
Rosenfeld Anatoly B.,
Zavgorodni Sergei F.
Publication year - 2015
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v16i5.5589
Subject(s) - imaging phantom , materials science , dosimeter , dosimetry , monte carlo method , slab , nuclear medicine , optics , physics , medicine , mathematics , statistics , geophysics
The purpose of this study is to compare performance of several dosimetric methods in heterogeneous phantoms irradiated by 6 and 18 MV beams. Monte Carlo (MC) calculations were used, along with two versions of Acuros XB, anisotropic analytical algorithm (AAA), EBT2 film, and MOS kin dosimeters. Percent depth doses (PDD) were calculated and measured in three heterogeneous phantoms. The first two phantoms were a 30 × 30 × 30   cm 3solid‐water slab that had an air‐gap of 20 × 2.5 × 2.35   cm 3 . The third phantom consisted of 30 × 30 × 5   cm 3solid water slabs, two 30 × 30 × 5   cm 3slabs of lung, and one 30 × 30 × 1   cm 3solid water slab. Acuros XB, AAA, and MC calculations were within 1% in the regions with particle equilibrium. At media interfaces and buildup regions, differences between Acuros XB and MC were in the range of + 4.4 % to − 12.8 % . MOS kin and EBT2 measurements agreed to MC calculations within ∼ 2.5 % , except for the first centimeter of buildup where differences of 4.5% were observed. AAA did not predict the backscatter dose from the high‐density heterogeneity. For the third, multilayer lung phantom, 6 MV beam PDDs calculated by all TPS algorithms were within 2% of MC. 18 MV PDDs calculated by two versions of Acuros XB and AAA differed from MC by up to 2.8%, 3.2%, and 6.8%, respectively. MOS kin and EBT2 each differed from MC by up to 2.9% and 2.5% for the 6 MV, and by − 3.1 % and ∼ 2 % for the 18 MV beams. All dosimetric techniques, except AAA, agreed within 3% in the regions with particle equilibrium. Differences between the dosimetric techniques were larger for the 18 MV than the 6 MV beam. MOS kin and EBT2 measurements were in a better agreement with MC than Acuros XB calculations at the interfaces, and they were in a better agreement to each other than to MC. The latter is due to their thinner detection layers compared to MC voxel sizes. PACS numbers: 87.55.K‐, 87.55.kd, 87.55.km, 87.53.Bn, 87.55.k

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here