z-logo
open-access-imgOpen Access
Development and clinical evaluation of a simple optical method to detect and measure patient external motion
Author(s) -
Barbés Benigno,
Azcona Juan Diego,
Prieto Elena,
Foronda José Manuel,
García Marina,
Burguete Javier
Publication year - 2015
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v16i5.5524
Subject(s) - imaging phantom , position (finance) , motion (physics) , trajectory , calibration , track (disk drive) , measure (data warehouse) , computer vision , computer science , amplitude , linear motion , artificial intelligence , physics , optics , finance , quantum mechanics , astronomy , database , economics , operating system
A simple and independent system to detect and measure the position of a number of points in space was devised and implemented. Its application aimed to detect patient motion during radiotherapy treatments, alert of out‐of‐tolerances motion, and record the trajectories for subsequent studies. The system obtains the 3D position of points in space, through its projections in 2D images recorded by two cameras. It tracks black dots on a white sticker placed on the surface of the moving object. The system was tested with linear displacements of a phantom, circular trajectories of a rotating disk, oscillations of an in‐house phantom, and oscillations of a 4D phantom. It was also used to track 461 trajectories of points on the surface of patients during their radiotherapy treatments. Trajectories of several points were reproduced with accuracy better than 0.3 mm in the three spatial directions. The system was able to follow periodic motion with amplitudes lower than 0.5 mm, to follow trajectories of rotating points at speeds up to 11.5 cm/s, and to track accurately the motion of a respiratory phantom. The technique has been used to track the motion of patients during radiotherapy and to analyze that motion. The method is flexible. Its installation and calibration are simple and quick. It is easy to use and can be implemented at a very affordable price. Data collection does not involve any discomfort to the patient and does not delay the treatment, so the system can be used routinely in all treatments. It has an accuracy similar to that of other, more sophisticated, commercially available systems. It is suitable to implement a gating system or any other application requiring motion detection, such as 4D CT, MRI or PET. PACS numbers: 87.55.N, 87.56.Da

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here