A dosimetry technique for measuring kilovoltage cone‐beam CT dose on a linear accelerator using radiotherapy equipment
Author(s) -
Scandurra Daniel,
Lawford Catherine E.
Publication year - 2014
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v15i4.4658
Subject(s) - dosimetry , imaging phantom , nuclear medicine , collimator , ionization chamber , linear particle accelerator , dose profile , radiation therapy , image guided radiation therapy , cone beam computed tomography , percentage depth dose curve , medicine , beam (structure) , physics , optics , radiology , computed tomography , ion , quantum mechanics , ionization
This work develops a technique for kilovoltage cone‐beam CT (CBCT) dosimetry that incorporates both point dose and integral dose in the form of dose length product, and uses readily available radiotherapy equipment. The dose from imaging protocols for a range of imaging parameters and treatment sites was evaluated. Conventional CT dosimetry using 100 mm long pencil chambers has been shown to be inadequate for the large fields in CBCT and has been replaced in this work by a combination of point dose and integral dose. Absolute dose measurements were made with a small volume ion chamber at the central slice of a radiotherapy phantom. Beam profiles were measured using a linear diode array large enough to capture the entire imaging field. These profiles were normalized to absolute dose to form dose line integrals, which were then weighted with radial depth to form the DLP CBCT . This metric is analogous to the standard dose length product (DLP), but derived differently to suit the unique properties of CBCT. Imaging protocols for head and neck, chest, and prostate sites delivered absolute doses of 0.9, 2.2, and 2.9 cGy to the center of the phantom, and DLP CBCT of 28.2, 665.1, and 565.3 mGy.cm, respectively. Results are displayed as dose per 100 mAs and as a function of key imaging parameters such as kVp, mAs, and collimator selection in a summary table. DLP CBCT was found to correlate closely with the dimension of the imaging region and provided a good indication of integral dose. It is important to assess integral dose when determining radiation doses to patients using CBCT. By incorporating measured beam profiles and DLP, this technique provides a CBCT dosimetry in radiotherapy phantoms and allows the prediction of imaging dose for new CBCT protocols. PACS number: 87.57.uq

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here