Homogeneity of GAFCHROMIC EBT2 film among different lot numbers
Author(s) -
Mizuno Hirokazu,
Takahashi Yutaka,
Tanaka Atsushi,
Hirayama Takamitsu,
Yamaguchi Tsuyoshi,
Katou Hiroaki,
Takahara Keiko,
Okamoto Yoshiaki,
Teshima Teruki
Publication year - 2012
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v13i4.3763
Subject(s) - homogeneity (statistics) , materials science , scanner , calibration curve , quality assurance , optics , nuclear medicine , statistics , mathematics , medicine , physics , external quality assessment , pathology , detection limit
EBT2 film is widely used for quality assurance in radiation therapy. The purpose of this study was to investigate the homogeneity of EBT2 film among various lots, and the dose dependence of heterogeneity. EBT2 film was positioned in the center of a flatbed scanner and scanned in transmission mode at 75 dpi. Homogeneity was investigated by evaluating gray value and net optical density (netOD) with the red color channel. The dose dependence of heterogeneity in a single sheet from five lots was investigated at 0.5, 2, and 3 Gy. Maximum coefficient of variation as evaluated by netOD in a single film was 3.0% in one lot, but no higher than 0.5% in other lots. Dose dependence of heterogeneity was observed on evaluation by gray value but not on evaluation by netOD. These results suggest that EBT2 should be examined in each lot number before clinical use, and that the dose calibration curve should be constructed using netOD. PACS number: 87

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here