Antiscatter grid use in pediatric digital tomosynthesis imaging
Author(s) -
King Jenna M.,
Elbakri Idris A.,
Reed Martin
Publication year - 2011
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v12i4.3641
Subject(s) - tomosynthesis , image quality , grid , radiography , medical physics , computer science , digital radiography , nuclear medicine , artificial intelligence , medicine , radiology , image (mathematics) , mammography , mathematics , geometry , cancer , breast cancer
The objective of this study was to assess the effect of antiscatter grid use on tomosynthesis image quality. We performed an observer study that rated the image quality of digital tomosynthesis scout radiographs and slice images of a Leeds TO.20 contrast‐detail test object embedded in acrylic with and without a grid. We considered 10, 15, 20 and 25 cm of acrylic to represent the wide range of patient thicknesses encountered in pediatric imaging. We also acquired and rated images without a grid at an increased patient dose. The readers counted the total number of visible details in each image as a measure of relative image quality. We observed that the antiscatter grid improves tomosynthesis image quality compared to the grid‐out case, which received image quality scores similar to grid‐in radiography. Our results suggest that, in order to achieve the best image quality in exchange for the increase in patient dose, it may often be appropriate to include an antiscatter grid for pediatric tomosynthesis imaging, particularly if the patient thickness is greater than 10 cm. PACS number: 87.57.‐s