z-logo
Premium
Anisotropy functions for 125 I and 103 Pd sources
Author(s) -
Weaver Keith
Publication year - 1998
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.598458
Subject(s) - anisotropy , monte carlo method , brachytherapy , computational physics , dosimetry , experimental data , physics , photon , optics , statistics , mathematics , nuclear medicine , radiation therapy , medicine
The anisotropy function was defined by AAPM Task Group 43 in a recent report, and reference data were provided. However, some of the data appeared to have large experimental errors. In this work, anisotropy functions are presented for three low‐energy photon‐emitting sources commonly used for brachytherapy:125 I source models 6702 and 6711, and the model 200103 Pd source. The data were generated via a two‐step process: the sources' intrinsic radiation emission pattern was determined from in‐air measurements, and these data were used as input to Monte Carlo calculations of the fluence distribution in water. The new anisotropy function values for iodine compare well with earlier data from diode measurements, and seem more physically realistic than some data provided by TG43. Previous palladium data measured with TLDs are more variable, but on the average are similar to the data presented here. The new data have estimated uncertainties of 3%–6%; this level of accuracy should be completely adequate for clinical calculations.

This content is not available in your region!

Continue researching here.