z-logo
Premium
Evidence for a significantly higher than expected depth dependence of wedge transmission factors on the 4 MV beam of a new dual energy accelerator
Author(s) -
Biggs P. J.,
Beatty J.,
Radloff J.
Publication year - 1998
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.598423
Subject(s) - wedge (geometry) , linear particle accelerator , beam (structure) , field size , optics , particle accelerator , transmission (telecommunications) , physics , beam energy , computational physics , electrical engineering , engineering
Wedge transmission factors have been measured for two sets of physical wedges for the 4 MV beam of a new dual energy linear accelerator as a function of field size and depth. The field size dependence of these factors has been compared with the 4 MV beam from another, single‐energy linear accelerator and a difference of a factor of approximately 2 has been observed in the relative wedge transmission factors between the two machines.