A Monte Carlo investigation of electron‐beam applicator scatter
Author(s) -
Ebert M. A.,
Hoban P. W.
Publication year - 1995
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.597414
Subject(s) - monte carlo method , electron , beam (structure) , fluence , scattering , cathode ray , collimated light , optics , secondary electrons , physics , component (thermodynamics) , materials science , nuclear physics , laser , statistics , mathematics , thermodynamics
An EGS4 Monte Carlo investigation into applicator scatter in clinical electron beams has been undertaken in order to establish the characteristics of electrons incident on the patient surface which have interacted with collimation systems. The applicator scattered component of an electron beam (including that from irregularly shaped cutouts) should be considered when modeling the electron phase space since it represents a component of the beam incident on the patient with widely varying characteristics to those of the primary beam. Scattering off an edge of applicator material is considered in terms of the types and characteristics of incident primary beam, the resulting interactions in the edge, and the fluence and energy characteristics of the emerging particles. Results indicate that the principal component to consider is scattered electrons due to the electron component of the primary beam, and that the fluence and energy characteristics of this component are dependent upon primary beam energy and the configuration of the applicator apertures.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here