X‐ray tube potential, filtration, and detector considerations in dual‐energy chest radiography
Author(s) -
Gauntt David M.,
Barnes Gary T.
Publication year - 1994
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.597373
Subject(s) - flat panel detector , noise (video) , detector , filtration (mathematics) , tube (container) , materials science , filter (signal processing) , radiography , x ray tube , mediastinum , nuclear medicine , biomedical engineering , physics , optics , radiology , medicine , mathematics , computer science , composite material , statistics , electrode , anode , quantum mechanics , artificial intelligence , image (mathematics) , computer vision
The effect of x‐ray tube potential and prepatient and interdetector filtration in single exposure dual energy chest imaging has been studied employing a carefully benchmarked model. The analysis utilized published methodology. Noise in simulated lung and mediastinum fields of the aluminum (bone) and Lucite (soft tissue) images were studied at fixed entrance skin exposure (ESE) for commonly employed sandwich detector and sandwich imaging plate configurations. Our results indicate noise in the lung increases slowly with tube potential above 120 kVp, while noise in the mediastinum decreases rapidly. Also, at high tube potential (≥120 kVp) adding moderate amounts of prepatient K ‐edge filtration (≅100 mg/cm 2 ) while optimizing imaging conditions for the lung tends to decrease noise in the lungs by ≅30% while increasing noise in the mediastinum by a similar amount. Without K ‐edge prepatient filtration, noise in the lung is minimized with Cu interdetector filter weights near 400 mg/cm 2 . In the mediastinum noise is minimized with heavier interdetector filter and prepatient K ‐edge filter weights. Prepatient K ‐edge filter weights that minimize image noise in either field can increase the tube loading by factors ranging from 10 to 10 10 . Systems designed with sandwich detectors using commercially available phosphors and coating weights can produce contrast‐to‐noise ratios (CNRs) as high as 50% of the theoretical limit (defined as an optimized system with a totally absorbing rear detector).

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here