z-logo
Premium
Determination of x‐ray tube focal spot position
Author(s) -
Kubota Hideaki,
Ozaki Yoshiaki,
Matsumoto Masao,
Kanamori Hitoshi
Publication year - 1993
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.597122
Subject(s) - x ray tube , position (finance) , optics , computed radiography , tube (container) , physics , x ray , materials science , medical physics , computer science , computer vision , image quality , electrode , quantum mechanics , composite material , anode , finance , economics , image (mathematics)
In x‐ray spectral measurement using a semiconductor detector system, accurate positioning of the detector in front of the anode is very important. This needs accurate determination of the focal spot position on the target in an x‐ray tube. We present a simple technique to determine the focal spot position accurately. Using a pinhole technique, the focal image made by x rays is overlapped on the target image made by visible rays emitted when filament heating takes places. Both images are made in the same geometry with the tube diaphragm and filter removed.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here