Recommendations on performance characteristics of diagnostic exposure meters: Report of AAPM Diagnostic X‐Ray Imaging Task Group No. 6
Author(s) -
Wagner L. K.,
Fontenla Doracy P.,
KimmeSmith Carolyn,
Rothenberg Lawrence N.,
Shepard Jeff,
Boone John M.
Publication year - 1992
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.596904
Subject(s) - task group , medical physics , radiography , calibration , medical imaging , nuclear medicine , mammography , medicine , radiation exposure , digital radiography , computed radiography , radiology , computer science , physics , engineering , artificial intelligence , image quality , engineering management , cancer , breast cancer , image (mathematics) , quantum mechanics
Task Group 6 of the Diagnostic X‐Ray Imaging Committee of the American Association of Physicists in Medicine (AAPM) was appointed to develop performance standards for diagnostic x‐ray exposure meters. The recommendations as approved by the Diagnostic X‐Ray Imaging Committee and the Science Council of the AAPM are delineated in this report and provide specifications on meter precision, calibration accuracy, calibration reference points, linearity, energy dependence, exposure rate dependence, leakage, amplification gain settings, directional dependence, the stem effect, constancy checks, and calibration intervals. The report summarizes recommendations for meters used in mammography, general purpose radiography including special procedures, computed tomography, and radiation safety surveys for x‐ray radiography.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here