Dual‐energy x‐ray projection imaging: Two sampling schemes for the correction of scattered radiation
Author(s) -
Wagner Frederick C.,
Macovski Albert,
Nishimura Dwight C.
Publication year - 1988
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.596188
Subject(s) - imaging phantom , optics , energy (signal processing) , physics , sampling (signal processing) , subtraction , image subtraction , projection (relational algebra) , image quality , contrast (vision) , fluoroscopy , signal to noise ratio (imaging) , scattering , iterative reconstruction , image processing , computer science , mathematics , algorithm , detector , computer vision , image (mathematics) , binary image , arithmetic , quantum mechanics , nuclear physics
In addition to the familiar problems of reduced contrast and signal‐to‐noise ratio (SNR) in the single energy case, dual‐energy subtractions in the presence of scattered radiation suffer further degradations from: (1) artifacts due to nonuniform subtraction of scatter, and (2) a serious deterioration of the signal of interest. To determine the expected performance of scatter correcting schemes, we simulated energy subtractions performed in the presence of scatter. We discuss scatter's detrimental effects on contrast and SNR in these simulations and the expected improvements from scatter corrections to within 5% to 10%. We introduce two sampling schemes for the correction of scatter. Each scheme requires two measurements, and each involves placing an x‐ray opaque sampling grid between the source and the object. In the first method, the grid is an array of lead disks present only during one measurement. Using these samples we generate an estimate of the scatter field and then subtract it from the second measurement yielding a scatter corrected image. In the second method, the grid is an array of lead strips present during both measurements but displaced between measurements by one‐half of a strip spacing to completely sample the image. From the two measurements we generate an image to be corrected, an estimate of the scatter field, and a scatter corrected image. In phantom studies implemented on a digital fluoroscopy system, we observed for single energy images of blood vessel phantoms improved contrast and field uniformity. For scatter corrected selective material cancellations in human phantoms we observed improved contrast and significant reduction in artifacts. In both cases we observed no significant loss in SNR. These results facilitate the implementation of efficient large area detectors for dual‐energy imaging.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here