Author(s) -
Siddon Robert L.,
Dewyngaert J. Keith,
Bjärngard Bengt E.
Publication year - 1985
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.595710
Subject(s) - square (algebra) , field (mathematics) , table (database) , point (geometry) , mathematics , dosimetry , geometry , mathematical analysis , computer science , nuclear medicine , pure mathematics , medicine , data mining
The concept of the equivalent field is used extensively in radiotherapy dose calculation algorithms. The rationale for using equivalent fields is to allow dose calculations for a wide variety of field shapes, while maintaining dose calculational data for only a few, very regularly shaped fields. A common example is the table of equivalent squares of rectangular fields presented by Day in the British Journal of Radiology . Recently, in searching for fast dose calculation algorithms for irregular fields, we introduced the concept of the equivalent square of a right triangular field. It is shown that an arbitrary irregular field of N vertices may be decomposed into 2 N right triangular fields, each with a precalculated equivalent square. The scatter at the point of calculation due to the irregular field is then obtained as a sum of the scatter contributions from the equivalent squares. The scatter integration with right triangles is compared with scatter integration using program irreg .