Computer dosimetry of 1 9 2 Ir wire
Author(s) -
Kline Robert W.,
Gillin Michael T.,
Grimm Daniel F.,
NiroomandRad Azam
Publication year - 1985
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.595685
Subject(s) - dosimetry , calibration , nuclear medicine , medical physics , computer program , radiation treatment planning , computer science , physics , radiation therapy , medicine , operating system , quantum mechanics
The dosimetry of 1 9 2 Ir linear sources with a commercial treatment planning computer system has been evaluated. Reference dose rate data were selected from the literature and normalized in a manner consistent with our clinical and dosimetric terminology. The results of the computer calculations are compared to the reference data and good agreement is shown at distances within about 7 cm from a linear source. The methodology of translating source calibration in terms of exposure rate for use in the treatment planning computer is developed. This may be useful as a practical guideline for users of similar computer calculation programs for iridium as well as other sources.