Dosimetric characterization of the 18‐MV photon beam from the Siemens Mevatron 77 linear accelerator
Author(s) -
Palta Jatinder R.,
Meyer Jeffrey A.,
Hogstrom Kenneth R.
Publication year - 1984
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.595564
Subject(s) - linear particle accelerator , collimator , ionization chamber , percentage depth dose curve , dosimetry , attenuation , beam (structure) , field size , linear regression , nuclear medicine , optics , photon , physics , materials science , mathematics , ionization , statistics , medicine , ion , quantum mechanics
A comprehensive set of dosimetric measurements has been made on the Mevatron 77.80.67 18‐MV photon beam. Percentage depth dose, dose in the buildup region, field size dependence of output, transmission through lead, tray attenuation, and isodose curves for the open and wedged fields were measured using an ionization chamber in water and polystyrene phantoms. These dosimetric measurements sufficiently characterized the beam to permit clinical use. The depth dose at 10‐cm depth for a 10×10 cm 2 field at 100‐cm source‐to‐skin distance (SSD) is 80.9%, which meets design specifications. Central axis depth‐dose data were fitted to within 0.5% by a set of polynomial equations utilizing a two‐dimensional linear regression analysis. Tissue–maximum ratios calculated from depth‐dose data agree with measured data to within 2%. Output differences as large as 2.5% were measured for rectangular fields depending on which collimator jaws defined the long dimension of the field. The field size dependence of output was fit to within ±0.1% by a linear regression. The half‐value thickness of the beam was measured to be 13 mm of lead.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here