The use of the ferrous sulphate dosimeter for intercomparison of absorbed dose from electron beams
Author(s) -
Chen W. L.,
Chang S. C.
Publication year - 1984
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.595509
Subject(s) - dosimeter , dosimetry , absorbed dose , calibration , radiochemistry , irradiation , radiation , nuclear medicine , materials science , environmental science , physics , nuclear physics , chemistry , medicine , quantum mechanics
An intercomparison of absorbed dose from high‐energy electron beams using Fricke dosimetry at four radiotherapy centers in the north of Taiwan is described. Two water phantoms of 30×30×30 cm 3 and a number of Fricke dosimeters were used at each center. The dosimeters were irradiated to an absorbed dose of approximately 45 Gy (4500 rad) at different electron energies. The measurements, calibration procedures, and the factors affecting the response of the dosimeters are discussed. The results of the intercomparison of the four centers are given in detail.