Scattered radiation in chest radiography
Author(s) -
Niklason Loren T.,
Sorenson James A.,
Nelson James A.
Publication year - 1981
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.595027
Subject(s) - radiography , mediastinum , computed radiography , contrast (vision) , nuclear medicine , radiology , fluoroscopy , x ray , materials science , medicine , optics , physics , image quality , computer science , artificial intelligence , image (mathematics)
Luminance scatter fractions were measured for patients and phantoms in various regions of the chest film. Scatter fractions were found to be highly variable, being quite high in the regions of the chest with a large equivalent tissue thickness (e.g. mediastinum, chest wall, and subdiaphramatic areas). For example, approximately 90% of the radiation reaching the mediastinum region of the chest film was found to be scattered radiation when a grid was not used. Even with a 12:1 grid, less than 50% of the available primary beam contrast was being displayed in the mediastinum. In regions of the chest film with high scatter fractions, film latitude was also a problem, further reducing image contrast in these regions. Image contrast and scatter fractions in chest radiography were investigated for different scatter rejection techniques including air gaps, antiscatter grids, and slit radiography.