Computer‐assisted quality assurance for radiographic equipment
Author(s) -
Glaze Sharon,
Schneiders Nicholas,
Archer Benjamin,
Bushong Stewart
Publication year - 1980
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.594698
Subject(s) - quality assurance , reproducibility , calibration , timer , radiography , medical physics , computer science , field size , dosimetry , nuclear medicine , medicine , computer hardware , mathematics , statistics , radiology , microcontroller , external quality assessment , pathology
A quality assurance procedure which includes a computer program has been implemented to facilitate measurement of some of the parameters normally tested on diagnostic radiographic equipment. These include: reproducibility of output, linearity of mA settings, filtration, kVp calibration, timer error, SID indication, light field—x‐ray field coincidence and effective focal spot size. The user may omit those sections which are not applicable to his needs. Full use is made of the data reducing the number of field measurements needed.