Characterization of Clinac‐18 electron‐beam energy using a magnetic analysis method
Author(s) -
Wessels B. W.,
Paliwal B. R.,
Parrott M. J.,
Choi M. C.
Publication year - 1979
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.594550
Subject(s) - characterization (materials science) , cathode ray , electron , beam (structure) , physics , atomic physics , materials science , nuclear magnetic resonance , nuclear physics , optics
Clinac‐18 electron‐beam energies and spectral energy distributions were measured by a magnetic analysis method to verify results obtained by the standard range method. A comparison of these measurements with nominal energies quoted by the manufacturer indicates that electron‐beam energies measured by both methods are in agreement with the manufacturer's specifications to ±0.6 MeV. Spectral energy distribution of the electron beam was measured to be approximately 1 MeV (FWHM) at all energies.