Premium
Victoreen calibration “discrepancies” noted
Author(s) -
Gray Laurence
Publication year - 1975
Publication title -
medical physics
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.594172
Subject(s) - calibration , condenser (optics) , ionization chamber , ionization , physics , medical physics , nuclear medicine , optics , mathematics , statistics , medicine , ion , light source , quantum mechanics
Errors of 12% are noted for condenser R‐Meters calibrated at the Victoreen Laboratory. Radiotherapists relying on these factors are cautioned to verify the calibration of each ionization chamber.