z-logo
Premium
Victoreen calibration “discrepancies” noted
Author(s) -
Gray Laurence
Publication year - 1975
Publication title -
medical physics
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.594172
Subject(s) - calibration , condenser (optics) , ionization chamber , ionization , physics , medical physics , nuclear medicine , optics , mathematics , statistics , medicine , ion , light source , quantum mechanics
Errors of 12% are noted for condenser R‐Meters calibrated at the Victoreen Laboratory. Radiotherapists relying on these factors are cautioned to verify the calibration of each ionization chamber.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here