Poster ‐ 22: Retrospective analysis of portal dosimetry based QA of Prostate VMAT Plans
Author(s) -
Badu Shyam,
Darko Johnson,
Fleck Andre,
Osei Ernest
Publication year - 2016
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4961796
Subject(s) - truebeam , quality assurance , dosimetry , nuclear medicine , medicine , linear particle accelerator , medical physics , beam (structure) , physics , optics , pathology , external quality assessment
Purpose: The purpose of this study is to retrospectively analyze the portal dosimetry based quality assurance of prostate VMAT plans. Methods: Our standard quality assurance of VMAT treatment plans are performed using EPID installed on Varian TrueBeam Linac. In our current study we analyzed 84 prostate pretreatment VMAT plans. All plans consisted of two arcs, 7800cGy in 39 fractions with a 6MV beam. For each of these VMAT plans, the measured fluence for each arc is compared with the reference fluence using gamma index analysis. Results: We have compared the gamma passing rates for three criteria; 3%/3mm, 2%/2mm and 1%/1mm. Out of 168 arcs measured, the number below the gamma passing rate 95% using the area, Field+1cm, are 0, 2, and 124 for 3%/3mm, 2%/2mm and 1%/1mm criteria respectively. Corresponding numbers for MLC CIAO are 0, 2, and 139 respectively. The average gamma passing rate for all arcs measured using Field+1cm are 99.9±0.4, 99.6±1.2, and 90.9±6.5 for 3%/3mm, 2%/2mm and 1%/1mm respectively. Similarly if the MLC CIAO area is analyzed, a passing rate of 99.9±0.2, 99.2±1.2 and 87.2±8.5 respectively was observed. The average of the maximum gamma was also found to increase with tighter criteria. Conclusion: Analysis of prostate VMAT quality assurance plans indicate that the gamma passing rate is sensitive to the criteria and the area analyzed.