Poster ‐ 07: Investigations of the Advanced Collapsed‐cone Engine for HDR Brachytherapy Scalp Treatments
Author(s) -
CawstonGrant Brie,
Morrison Hali,
Sloboda Ron,
Me Geetha
Publication year - 2016
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4961781
Subject(s) - imaging phantom , nuclear medicine , brachytherapy , materials science , skull , biomedical engineering , dosimetry , medicine , radiation therapy , radiology , surgery
Purpose: To present an investigation of the Advanced Collapsed‐cone Engine (ACE) in Oncentraê Brachy (OcB) v4.5 using a tissue equivalent phantom modeling scalp brachytherapy (BT) treatments. Methods: A slab phantom modeling the skin, skull, brain and mold was used. A dose of 400cGy was prescribed to just above the skull layer using TG‐43 and was delivered using an HDR afterloader. Measurements were made using Gafchromic™ EBT3 film at four depths within the phantom. The TG‐43 planned and film measured doses were compared to the standard (sACE) and high (hACE) accuracy ACE options in OcB between the surface and below the skull. Results: The average difference between the TG‐43 calculated and film measured doses was −11.25±3.38% when there was no air gap between the mold and skin; sACE and hACE doses were on average lower than TG‐43 calculated doses by 3.41±0.03% and 2.45±0.03%, respectively. With a 3mm air gap between the mold and skin, the difference between the TG‐43 calculated and measured doses was −8.28±5.76%; sACE and hACE calculations yielded average doses 1.87±0.03% and 1.78±0.04% greater than TG‐43, respectively. Conclusions: TG‐43, sACE, and hACE were found to overestimate doses below the skull layer compared to film. With a 3mm air gap between the mold and skin, sACE and hACE more accurately predicted the film dose to the skin surface than TG‐43. More clinical variations and their implications are currently being investigated.