TH‐CD‐BRA‐04: Effect of a Strong Magnetic Field On TLDs, OSLDs, and Gafchromic Films Using An Electromagnet
Author(s) -
Wang J,
Rubinstein A,
Ohrt J,
Ibbott G,
Wen Z
Publication year - 2016
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4958147
Subject(s) - dosimeter , thermoluminescent dosimeter , electromagnet , irradiation , imaging phantom , materials science , optics , dosimetry , magnetic field , magnet , nuclear medicine , optically stimulated luminescence , thermoluminescence , physics , radiation , nuclear physics , medicine , quantum mechanics
Purpose: To study the effect of strong magnetic field on three types of dosimeters using an electromagnet inside a Linac vault. Materials and Methods: Three types of dosimeters, thermoluminescent Dosimeters (TLDs), optically stimulated luminescent Dosimeters (OSLDs), and EBT3 Film were used to measure radiation dose response inside an electromagnet that could produce a strong magnetic field (B>1.5 T). The dosimeters were placed inside a plastic phantom between the two poles of the magnet, at approximately 3 meters from the iso‐center of an Elekta Versa HD Linac. The B field was calibrated with a Gauss meter (Model: GM‐2, AlphaLab Inc). The dosimeters received ∼2 Gy with and without the presence of the 1.5 T magnetic field. The EBT3 films were scanned 24 hours before and 24 hours after irradiation. The TLD dosimeters were read 1 week after irradiation. The OSLDs were read two weeks after irradiation. The ratios of signals of dosimeters irradiated with the B field to the signals without the B field were calculated. Two experiments have been conducted so far. Results: The ratios (averaged over two experiments) of dosimeter signals with vs without B field were 0.994 for films, 0.994 for OSLDs, and 1.002 for TLDs. The statistical uncertainty was ∼3%. Conclusions: The three types of dosimeters (film, TLD, OSLD) seem not affected by the presence of a magnetic field (B=1.5 T) with the uncertainty of ∼3%. They may be suitable for QA purposes in a strong B field up to 1.5 T. More measurements will be conducted for reproducibility testing. We acknowledge research support from Elekta AB.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here