z-logo
Premium
SU‐F‐T‐135: A Retrospective Analysis of the Impact of Range Uncertainty in Brain Patients
Author(s) -
Grantham K,
Santanam L,
Goddu S,
Sun B,
Zhang T,
Mutic S,
Robinson C,
Huang J,
Perkins S,
Tsien C,
Bradley J,
Zhao T
Publication year - 2016
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4956271
Subject(s) - optic chiasm , medicine , optic nerve , retrospective cohort study , nuclear medicine , ophthalmology , surgery
Purpose: We retrospectively evaluate the dosimetric impact of a 3.5% range uncertainty on CTV coverage and normal organ toxicity for a cohort of brain patients. Methods: Twenty treatment plans involving 20 brain cancer patients treated with Mevions S250 were reviewed. Forty uncertain plans were made by changing the ranges in original plans by ±3.5% while keeping all devices unchanged. Fidelity to the original plans was evaluated with gamma index. Changes in generalized equivalent uniform dose (gEUD) were reported for the following structures: CTV coverage, brainstem, optic chiasm, and optic nerves. Comparisons were made by plotting the relevant endpoints from the uncertain plans as a function of the same endpoints from the original clinical plan. Results: Gamma‐index analysis resulted in a 50% pass rate of the uncertain plans using a 90% passing rate and 3%/3mm criterion. A 9.5% decrease in the slope of gEUD plot for the CTV was observed for the 3.5% downward range shift. However, the change in slope did not result in a gEUD change greater than 1.1% for the CTV. The slopes of the gEUD plots for normal structures increased by 3.1% 3.9% 2.4% and 0.2% for the chiasm, brainstem, left optic nerve and right optic nerve respectively. The maximum deviation from the gEUD of the clinical plan for normal structures was: 64% in the chiasm, 31% for the brainstem, and 19% for both optic nerves. Conclusion: A retrospective review shows moderate radiobiological impact of range uncertainty in passively scattered proton therapy with sporadic catastrophe. The linear regression analysis on the statistical data indicates a systematic deviation of gEUD from treatment planning in the light of range uncertainty.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here