SU‐F‐P‐23: Setup Uncertainties for the Lung Stereotactic Body Radiation Therapy
Author(s) -
Zhang Q,
Cao Y,
Jamshidi A,
Vigneri P,
Madu C,
Potters L,
Klein E
Publication year - 2016
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4955730
Subject(s) - radiography , nuclear medicine , standard deviation , radiosurgery , position (finance) , radiation therapy , medical imaging , computer science , physics , optics , medicine , mathematics , radiology , artificial intelligence , statistics , finance , economics
Purpose: The Exactrack X‐ray system with six degree‐of‐freedom (6DoF) adjustment ability can be used for setup of lung stereotactic body radiation therapy. The setup uncertainties from ExacTrack 6D system were analyzed. Methods: The Exactrack X‐ray 6D image guided radiotherapy system is used in our clinic. The system is an integration of 2 subsystems: (1): an infrared based optical position system and (2) a radiography kV x‐ray imaging system. The infrared system monitors reflective body markers on the patient's skin to assistant in the initial setup. The radiographic kV devices were used for patient positions verification and adjustment. The position verification was made by fusing the radiographs with the digitally reconstructed radiograph (DRR) images generated by simulation CT images using 6DoF fusion algorithms. Those results were recorded in our system. Gaussian functions were used to fit the data. Results: For 37 lung SBRT patients, the image registration results for the initial setup by using surface markers and for the verifications, were measured. The results were analyzed for 143 treatments. The mean values for the lateral, longitudinal, vertical directions were 0.1, 0.3 and 0.3mm, respectively. The standard deviations for the lateral, longitudinal and vertical directions were 0.62, 0.78 and 0.75mm respectively. The mean values for the rotations around lateral, longitudinal and vertical directions were 0.1, 0.2 and 0.4 degrees respectively, with standard deviations of 0.36, 0.34, and 0.42 degrees. Conclusion: The setup uncertainties for the lung SBRT cases by using Exactrack 6D system were analyzed. The standard deviations of the setup errors were within 1mm for all three directions, and the standard deviations for rotations were within 0.5 degree.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here