TU‐G‐BRD‐05: Results From Multi‐Institutional Measurements with An Anthropomorphic Spine Phantom
Author(s) -
Molineu A,
Hernandez N,
Alvarez P,
Followill D
Publication year - 2015
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4925740
Subject(s) - thermoluminescent dosimeter , imaging phantom , nuclear medicine , medicine , pinnacle , radiation treatment planning , dicom , dosimetry , medical physics , radiation therapy , radiology , dosimeter
Purpose: To analyze the results from an anthropomorphic spine phantom used for credentialing institutions for National Cancer Institute (NCI) sponsored clinical trial. Methods: An anthropomorphic phantom that contains left and right lungs, a heart, an esophagus, spinal cord, bony material and a PTV was sent to institutions wishing to be credentialed for NCI trials. The PTV holds 4 TLD and radiochromic film in the axial and sagittal planes. The heart holds one TLD. Institutions created IMRT plans to cover ≥90% of the PTV with 6 Gy and limit the cord dose to <0.35cc receiving 3.75 Gy and <1.2cc receiving 2.63 Gy. They were instructed to treat the phantom as they would a patient, including making plan specific IMRT/SBRT QA measurements before treatment. The TLD results in the PTV were required to be within ±7% of the plan dose. A gamma calculation was performed using the film results and the submitted DICOM plan. ≥85% of the analyzed region was required to pass a 5%/3 mm criteria. Results: 176 institutions irradiated the spine phantom for a total of 255 results. The pass rate was 73% (187 irradiations) overall. 44 irradiations failed only the gamma criteria, 2 failed only the dose criteria and 22 failed both. The most used planning systems were Eclipse (116) and Pinnacle (52) and they had pass rates of 76% and 71%, respectively. The AAA algorithm had a pass rate of 77% while superposition type algorithms had a 71% pass rate. The average TLD measurement to institution calculation ratio was 0.99 (0.04 std dev.). The average percent pixels passing the gamma criteria for films was 89% (12% std dev.) Conclusion: Results show that this phantom is an important part of credentialing and that we have room for improvement in IMRT/SBRT spine treatments. This work was supported by PHS CA180803 and CA037422 awarded by NCI, DHHS.