SU‐E‐T‐437: Dosimetric Assessment of Brass Mesh Bolus for Postmastectomy Chest Wall Irradiation
Author(s) -
Manger R,
Paxton A,
Cervino L
Publication year - 2014
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4888770
Subject(s) - bolus (digestion) , imaging phantom , brass , irradiation , nuclear medicine , ionization chamber , dosimetry , materials science , medicine , biomedical engineering , fluoroscopy , radiology , surgery , ionization , physics , copper , ion , quantum mechanics , nuclear physics , metallurgy
Purpose: It has been suggested that the use of a brass mesh bolus for chest wall irradiation sufficiently increases surface dose while having little effect on the dose at depth. This work quantified the increase in surface dose when using a brass mesh bolus in postmastectomy chest wall radiotherapy compared to tissue‐equivalent bolus and assessed its effect on dose at depth. Methods: Percent depth doses with brass bolus, 5mm tissue‐equivalent bolus, and no bolus were determined for a 6 MV photon beam in a solid water phantom using a parallel plate ionization chamber. Gafchromic film was used to determine the surface dose for the same three experimental setups. For comparison to a realistic treatment setup, gafchromic film and OSLDs were used to determine the surface dose over the irradiated area of a 6 MV chest wall plan with tangential beams delivered to a heterogeneous thorax phantom. The plan was generated using a CT of the phantom and delivered using brass mesh bolus, 5mm tissue‐equivalent bolus, and no bolus. Results: For the en face beam, the central surface dose increased to 90% of maximum with the tissue‐equivalent bolus, but to only 62% of maximum with the brass mesh. Using tangential beams on the thorax phantom, the surface dose increased from 40–72% to 75–110% of prescribed dose, with the brass mesh, and to 85–109% with the tissue‐equivalent bolus. At depths beyond dmax in the plastic water phantom, the dose with and without brass mesh bolus differed by less than 0.5%. Conclusion: A brass mesh may be considered as a substitute for tissue‐equivalent bolus to increase the superficial dose of 6 MV chest wall tangent plans. The brass mesh does not significantly change the dose at depth, so a non‐bolus plan could be used for bolus and non‐bolus treatments.