SU‐E‐T‐437: Dosimetric Assessment of Brass Mesh Bolus for Postmastectomy Chest Wall Irradiation
Author(s) -
Manger R,
Paxton A,
Cervino L
Publication year - 2014
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4888770
Subject(s) - bolus (digestion) , imaging phantom , brass , irradiation , nuclear medicine , ionization chamber , dosimetry , materials science , medicine , biomedical engineering , fluoroscopy , radiology , surgery , ionization , physics , copper , ion , quantum mechanics , nuclear physics , metallurgy
Purpose: It has been suggested that the use of a brass mesh bolus for chest wall irradiation sufficiently increases surface dose while having little effect on the dose at depth. This work quantified the increase in surface dose when using a brass mesh bolus in postmastectomy chest wall radiotherapy compared to tissue‐equivalent bolus and assessed its effect on dose at depth. Methods: Percent depth doses with brass bolus, 5mm tissue‐equivalent bolus, and no bolus were determined for a 6 MV photon beam in a solid water phantom using a parallel plate ionization chamber. Gafchromic film was used to determine the surface dose for the same three experimental setups. For comparison to a realistic treatment setup, gafchromic film and OSLDs were used to determine the surface dose over the irradiated area of a 6 MV chest wall plan with tangential beams delivered to a heterogeneous thorax phantom. The plan was generated using a CT of the phantom and delivered using brass mesh bolus, 5mm tissue‐equivalent bolus, and no bolus. Results: For the en face beam, the central surface dose increased to 90% of maximum with the tissue‐equivalent bolus, but to only 62% of maximum with the brass mesh. Using tangential beams on the thorax phantom, the surface dose increased from 40–72% to 75–110% of prescribed dose, with the brass mesh, and to 85–109% with the tissue‐equivalent bolus. At depths beyond dmax in the plastic water phantom, the dose with and without brass mesh bolus differed by less than 0.5%. Conclusion: A brass mesh may be considered as a substitute for tissue‐equivalent bolus to increase the superficial dose of 6 MV chest wall tangent plans. The brass mesh does not significantly change the dose at depth, so a non‐bolus plan could be used for bolus and non‐bolus treatments.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom