z-logo
Premium
SU‐E‐T‐265: Presage Thin Sheet Dosimeter Characterization
Author(s) -
Dumas M,
Rakowski J
Publication year - 2014
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4888596
Subject(s) - dosimeter , scanner , materials science , saturation (graph theory) , dosimetry , dose profile , nuclear medicine , analytical chemistry (journal) , optics , physics , chemistry , medicine , mathematics , chromatography , combinatorics
Purpose: To quantify the sensitivity and stability of the Presage dosimeter in sheet form for different concentrations of chemicals and for a diverse range of clinical photon energies. Methods: Presage polymer dosimeters are formulated to investigate and optimize their sensitivity and stability. The dosimeter is composed of clear polyurethane base, leucomalachite green reporting dye, and bromoform radical initiator in 1mm thick sheets. The chemicals are well mixed together, cast in an aluminum mold, and left to cure at 60 psi for a minimum of 2 days. Dosimeter response will be characterized at multiple energies including Co‐60, 6 MV, 15 MV, 50 kVp, and 250 kVp. The dosimeters are read by an Epson 10000 XL scanner at 800 dpi, 2^16 bit depth. Red component images are analyzed with ImageJ. Results: Analysis of optical density verse dose for Co‐60 energies indicates that the bromoform containing Presage was able to quantify dose from 0 to 300 Gy, with saturation beyond 300 Gy. Initial results show two regions of linear response, 0–100 Gy and 150–300 Gy. The 150–300 Gy region has a sensitivity of 0.0024 net OD/Gy. Further results on other energies are still in progress. Conclusions: This work shows the potential for use of thin sheets of Presage dosimeter as a dosimeter capable of being analyzed with a flatbed scanner.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here