SU‐E‐J‐38: Comparison of 6MV Photon Dose in a Perpendicular and Parallel Magnetic Field
Author(s) -
Ghila A,
Fallone B,
Rathee S
Publication year - 2014
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4888090
Subject(s) - imaging phantom , magnetic field , electromagnet , transverse plane , air gap (plumbing) , magnet , solenoid , electron , linear particle accelerator , physics , materials science , cathode ray , optics , nuclear magnetic resonance , beam (structure) , nuclear physics , structural engineering , quantum mechanics , composite material , engineering
Purpose: Integrating a linac with an MRI system would allow for real time tumour tracking however the patient will be irradiated in the presence of a magnetic field. The present study experimentally investigates the magnetic field effects on entrance, exit, and interface dose for both transverse and parallel magnetic fields. Methods: Polystyrene was used to construct a set of phantoms for Gafchromic film measurements. One phantom had an adjustable air gap and four other phantoms had one surface at various angles. The linac‐MR prototype consisting of a biplanar permanent magnet coupled to a linac was used for the transverse magnetic field measurements. A couple of solenoid electromagnets, stacked on top of each other and irradiated along their bore, were used for the parallel field measurements. Results: All doses are relative to no magnetic field. The transverse magnetic field reduced the entrance dose for all surface angles by strongly deflecting the contaminant electrons. The exit dose in a transverse magnetic field was found to be significantly higher. The entrance dose with a parallel magnetic field present is higher due to the contaminant electrons being concentrated within the beam area. The air gap phantom measurements, done in a transverse magnetic field, show a significant increase of the dose at the proximal side of the air gap and a decrease at the distal side. The measurements, done in the parallel magnetic field, show the concentration of secondary electrons in the air gap. Conclusion: The radiation dose measurements of a 6MV beam in a parallel and transverse magnetic field presented here are currently being replicated using Monte Carlo simulations. This verified Monte Carlo system could provide the dose calculation basis for future linac‐MR systems.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here