SU‐E‐J‐38: Comparison of 6MV Photon Dose in a Perpendicular and Parallel Magnetic Field
Author(s) -
Ghila A,
Fallone B,
Rathee S
Publication year - 2014
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4888090
Subject(s) - imaging phantom , magnetic field , electromagnet , transverse plane , air gap (plumbing) , magnet , solenoid , electron , linear particle accelerator , physics , materials science , cathode ray , optics , nuclear magnetic resonance , beam (structure) , nuclear physics , structural engineering , quantum mechanics , composite material , engineering
Purpose: Integrating a linac with an MRI system would allow for real time tumour tracking however the patient will be irradiated in the presence of a magnetic field. The present study experimentally investigates the magnetic field effects on entrance, exit, and interface dose for both transverse and parallel magnetic fields. Methods: Polystyrene was used to construct a set of phantoms for Gafchromic film measurements. One phantom had an adjustable air gap and four other phantoms had one surface at various angles. The linac‐MR prototype consisting of a biplanar permanent magnet coupled to a linac was used for the transverse magnetic field measurements. A couple of solenoid electromagnets, stacked on top of each other and irradiated along their bore, were used for the parallel field measurements. Results: All doses are relative to no magnetic field. The transverse magnetic field reduced the entrance dose for all surface angles by strongly deflecting the contaminant electrons. The exit dose in a transverse magnetic field was found to be significantly higher. The entrance dose with a parallel magnetic field present is higher due to the contaminant electrons being concentrated within the beam area. The air gap phantom measurements, done in a transverse magnetic field, show a significant increase of the dose at the proximal side of the air gap and a decrease at the distal side. The measurements, done in the parallel magnetic field, show the concentration of secondary electrons in the air gap. Conclusion: The radiation dose measurements of a 6MV beam in a parallel and transverse magnetic field presented here are currently being replicated using Monte Carlo simulations. This verified Monte Carlo system could provide the dose calculation basis for future linac‐MR systems.