Delivery validation of an automated modulated electron radiotherapy plan
Author(s) -
Connell T.,
Alexander A.,
Papaconstadopoulos P.,
Serban M.,
Devic S.,
Seuntjens J.
Publication year - 2014
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4876297
Subject(s) - collimator , collimated light , imaging phantom , radiation treatment planning , dosimetry , interfacing , nuclear medicine , monte carlo method , medical physics , radiation therapy , computer science , physics , medicine , optics , mathematics , radiology , computer hardware , laser , statistics
Purpose: Modulated electron radiation therapy (MERT) represents an active area of interest that offers the potential to improve healthy tissue sparing in treatment of certain cancer cases. Challenges remain however in accurate beamlet dose calculation, plan optimization, collimation method, and delivery accuracy. In this work, the authors investigate the accuracy and efficiency of an end‐to‐end MERT plan and automated delivery method. Methods: Treatment planning was initiated on a previously treated whole breast irradiation case including an electron boost. All dose calculations were performed using Monte Carlo methods and beam weights were determined using a research‐based treatment planning system capable of inverse optimization. The plan was delivered to radiochromic film placed in a water equivalent phantom for verification, using an automated motorized tertiary collimator. Results: The automated delivery, which covered four electron energies, 196 subfields, and 6183 total MU was completed in 25.8 min, including 6.2 min of beam‐on time. The remainder of the delivery time was spent on collimator leaf motion and the automated interfacing with the accelerator in service mode. Comparison of the planned and delivered film dose gave 3%/3mm gamma pass rates of 62.1%, 99.8%, 97.8%, 98.3%, and 98.7% for the 9, 12, 16, and 20 MeV, and combined energy deliveries, respectively. Delivery was also performed with a MapCHECK device and resulted in 3%/3  mm gamma pass rates of 88.8%, 86.1%, 89.4%, and 94.8% for the 9, 12, 16, and 20 MeV energies, respectively. Conclusions: Results of the authors’ study showed that an accurate delivery utilizing an add‐on tertiary electron collimator is possible using Monte Carlo calculated plans and inverse optimization, which brings MERT closer to becoming a viable option for physicians in treating superficial malignancies.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here