TU‐E‐108‐03: Optimization of the Beam‐On Latency of An Elekta Linac in the Delivery of Gated VMAT
Author(s) -
Cui G,
Housley D,
Chen F,
Mehta V,
Shepard D
Publication year - 2013
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4815418
Subject(s) - linear particle accelerator , gating , imaging phantom , nuclear medicine , medicine , dosimetry , beam (structure) , multileaf collimator , physics , optics , physiology
Purpose: To optimize the beam‐on latency of an Elekta linac in delivering gated VMAT. Methods: Step‐wise optimization was performed using a pre‐clinical gating interface developed by Elekta. Phantom studies were performed where surface motion was simulated using a moving chest plate. Two gating windows 77% and 66% around the end of exhalation were used for the gated VMAT deliveries. The treatment plans from three lung patients previously treated with un‐gated 6 MV SBRT were used for this evaluation. Baseline of the average beam‐on latency of the gated deliveries was first established. Then the average beam‐on latencies between a binned dose rate and a continuously variable dose rate (CVDR) deliveries were compared. With the CVDR delivery, further comparison between two different linac gun hold‐on times 1.35 s and 6.50 s was performed. The dosimetric accuracy of the gated deliveries was evaluated by comparing the measured with planned coronal dose distributions using gamma‐index analyses (3 mm/3% passing rate criteria). Results: The average beam‐on latencies of the gated deliveries with the CVDR were greatly reduced as compared with those of the binned dose rate for gating windows higher than 66%. With the CVDR deliveries, longer gun hold‐on times (6.50 s as compared to 1.35 s) made a dramatic improvement on the beam‐on latency for gating windows equal to and lower than 66%. There were insignificant dosimetric differences between all the gated deliveries and the un‐gated deliveries. Conclusion: For gated VMAT deliveries using an Elekta linac with the CVDR and the gun‐hold time of 6.50 s, the average beam‐on latency was reduced to a level (< 0.3 s) that has minimal impact on clinical results. High dosimetric accuracy was demonstrated for as many as 214 beam interruptions during a single 360‐degree‐arc delivery with gamma‐index passing rates were no lower than 99.0% for all tests. The research is partly supported by a grant from Elekta AB.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom