Monte Carlo modeling for dose assessment in cone beam CT for oral and maxillofacial applications
Author(s) -
Zhang G.,
Marshall N.,
Bogaerts R.,
Jacobs R.,
Bosmans H.
Publication year - 2013
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4810967
Subject(s) - monte carlo method , imaging phantom , cone beam computed tomography , dosimetry , medical imaging , nuclear medicine , beam (structure) , medical physics , medicine , physics , computed tomography , optics , mathematics , radiology , statistics
Purpose: To adjust Monte Carlo modeling for dose assessment in dedicated cone beam computed tomography (CBCT) of the oral and maxillofacial region.Methods: Two different CBCT systems with different fields of view (FOVs), beam qualities, and scan geometries were modeled using Monte Carlo simulation. Dose calculations for typical CBCT examinations were performed with the head and neck part of four computational anatomical phantoms.Results: Simulation results compared favorably to values acquired experimentally using physical phantoms (in the literature). For a given phantom scanned with 90 kV and 60 × 60 mm FOV, effective dose per mAs was on average the same for the two different systems. Exposing the four phantoms under identical settings for the same CBCT system resulted in variations in organ doses of greater than 100%, leading to differences in effective dose of 30%. For one system, the dose dependence on the operating tube potential can be described with a quadratic polynomial function. Dose distributions over the axial plane were presented as contour plots.Conclusions: Monte Carlo modeling is an efficient and accurate means of evaluating dose distributions for dedicated cone beam oral and maxillofacial CT. Results suggest large dose differences among patients undergoing the same examination on a given system, supporting approaches toward patient specific dosimetry. A dedicated and standardized computational phantom for head and neck dosimetry should be established.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here