Monte Carlo simulation based study of a proposed multileaf collimator for a telecobalt machine
Author(s) -
Sahani G.,
Dutt Sharma Sunil,
Dash Sharma P. K.,
Sharma D. N.,
Hussain S. A.
Publication year - 2013
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4773308
Subject(s) - monte carlo method , multileaf collimator , isocenter , imaging phantom , dosimetry , optics , flatness (cosmology) , beam (structure) , collimator , physics , computer science , materials science , linear particle accelerator , mathematics , nuclear medicine , statistics , medicine , cosmology , quantum mechanics
Purpose: The objective of the present work was to propose a design of a secondary multileaf collimator (MLC) for a telecobalt machine and optimize its design features through Monte Carlo simulation.Methods: The proposed MLC design consists of 72 leaves (36 leaf pairs) with additional jaws perpendicular to leaf motion having the capability of shaping a maximum square field size of 35 × 35 cm 2 . The projected widths at isocenter of each of the central 34 leaf pairs and 2 peripheral leaf pairs are 10 and 5 mm, respectively. The ends of the leaves and the x‐jaws were optimized to obtain acceptable values of dosimetric and leakage parameters. Monte Carlo N‐Particle code was used for generating beam profiles and depth dose curves and estimating the leakage radiation through the MLC. A water phantom of dimension 50 × 50 × 40 cm 3 with an array of voxels (4 × 0.3 × 0.6 cm 3 = 0.72 cm 3 ) was used for the study of dosimetric and leakage characteristics of the MLC. Output files generated for beam profiles were exported to the PTW radiation field analyzer software through locally developed software for analysis of beam profiles in order to evaluate radiation field width, beam flatness, symmetry, and beam penumbra.Results: The optimized version of the MLC can define radiation fields of up to 35 × 35 cm 2 within the prescribed tolerance values of 2 mm. The flatness and symmetry were found to be well within the acceptable tolerance value of 3%. The penumbra for a 10 × 10 cm 2 field size is 10.7 mm which is less than the generally acceptable value of 12 mm for a telecobalt machine. The maximum and average radiation leakage through the MLC were found to be 0.74% and 0.41% which are well below the International Electrotechnical Commission recommended tolerance values of 2% and 0.75%, respectively. The maximum leakage through the leaf ends in closed condition was observed to be 8.6% which is less than the values reported for other MLCs designed for medical linear accelerators.Conclusions: It is concluded that dosimetric parameters and the leakage radiation of the optimized secondary MLC design are well below their recommended tolerance values. The optimized design of the proposed MLC can be integrated into a telecobalt machine by replacing the existing adjustable secondary collimator for conformal radiotherapy treatment of cancer patients.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here