SU‐E‐T‐650: Impact of Machine Output Rate on Effective Dose Rate and Biologically Equivalent Fraction Size in Total Body Irradiation (TBI)
Author(s) -
Stewart RD,
Kalet A,
Laramore GE
Publication year - 2011
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3612613
Subject(s) - fraction (chemistry) , nuclear medicine , total body irradiation , materials science , medicine , biomedical engineering , chemistry , surgery , organic chemistry , chemotherapy , cyclophosphamide
Purpose: Examine the feasibility and biological significance of increasing machine output rates in extended‐distance Total Body Irradiation (TBI). Methods: Biologically equivalent doses for TBI are derived from the LQ cell survival model. Effective dose rates and corrected fraction size for a mini‐TBI (i.e., 1 fraction × 2 Gy to patient mid‐line) delivered using 4 and 8 beams are computed as a function of machine output rate. The number of beams used for a TBI tends to increase with increasing patient size because of the clinical constraint of no more than 1000 MU per beam. All of the reported results include a 1 minute gap between beams and a (conservative) 5 minute gap for patient repositioning. Results: For a 4 beam treatment, increasing the machine output rate from 200 to 400 MU/minute reduced the fraction delivery time by 36% (28 to 18 minutes), and only increased the effective dose rate from 7 to 11 cGy/minute. For 8 beams, the fraction delivery times for 200 and 400 MU/min are 52 minutes (3.85 cGy/min) to 20 minutes (6.16 cGy/min), respectively. Increasing the machine output rate from 200 to 400 MU/min is unlikely to require more than 2–5% reduction in fraction size (alpha/beta = 1 Gy). The size of the required correction decreases as alpha/beta increases. Conclusions: Patient‐specific planning to correct for differences in beam number (patient size) are unlikely to alter the determination of biologically equivalent fraction sizes by more than 5%, which is well within the +/− 10% uniformity specified in many TBI protocols. A clinical trial to evaluate patient safety and treatment efficacy for machine output rates as high as 600–700 MU per minute are reasonable. Such a trial has the potential to substantially to increase patient comfort and reduce facility and labor costs without jeopardizing treatment effectiveness or safety.