SU‐C‐BRA‐05: Commissioning of Rotational Total Skin Electron Irradiation on Trilogy Machine
Author(s) -
Zhang Q,
Chan R
Publication year - 2011
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3611465
Subject(s) - imaging phantom , beam (structure) , materials science , optics , linear particle accelerator , dosimetry , irradiation , nuclear medicine , physics , medicine , nuclear physics
Purpose: To commission 6MEV beam in HDTSe mode on a Varian Trilogy machine for rotational total skin treatment. Methods: HDTSe 6MEV beam with field size 40cm × 40cm at 380cm SSD was planned for total skin irradiation treatment. A custom designed flattening filter was added to improve beam uniformity. Rotational speed during treatment was to set at 3–5 revolution/minute. Beam calibration was at 100cm SSD following TG51 protocol. Beam parameters at extended distance were measured using GafChromic EBT2 films placed in solid water phantom. 30cm × 30cm square phantom and 33cm diameter round phantom were for stationary and rotational target respectively. Dose distribution during rotational treatment was verified by GafChromic films sandwiched inside a Rando phantom. Beam uniformity was verified using Kodak EDR2 ready pack films on a ½” thick, 4′×8′ plexiglas hanging perpendicular to the beam at distance. Results: At SSD of 380cm, with filter in place, stationary beam output was 89.6cGy/1000mu; dmax is at 0.9 cm; R50 is 1.82 cm and Rp is2.4cm. With rotational target, dmax is shifted to surface with R50 at 1.54 cm and output at surface was 38.8cGy/1000mu. Measured rotational factor was 0.433. X‐ray contamination is of the order of 3% for rotational target. Dose uniformity along the vertical axis from −90 cm to about 80cm is within +/− 5%, and within +/−10% from ‐95cm to 95cm. For the horizontal axis, the entire range is covered with the dose of better than 95%. Dose distribution using the rotational technique was shown to be uniform at different levels of a Rando phantom Conclusions: Dosimetric properties of the modified HDTSe 6MEV beam on a Varian Trilogy machine are verified to be optimal for rotational total skin irradiation treatment.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here