TU‐E‐BRB‐07: Optimal Treatment Beam Fluence Generation for Volumetric Arc Therapy Using Dose Image Backprojection with Initial Corona Calculation
Author(s) -
Barbiere J,
Hanley J
Publication year - 2010
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3469290
Subject(s) - imaging phantom , iterative reconstruction , iterative method , medical imaging , algorithm , fluence , computer science , mathematics , nuclear medicine , physics , computer vision , optics , artificial intelligence , medicine , laser
Purpose: Derive a novel solution to the initial premise that IMRT fluence can be easily calculated using dose image backprojection. Method and Materials: Calculations were performed using MATLAB with the imaging toolbox using parallel beam geometry. A phantom matrix P was constructed with a concave target, simultaneous integrated boost, and organ at risk with respective desired percent doses of 100, 110, and 40. Let R:SP be the backprojection of sinogram S of P. Iterative reconstruction using standard Ratio Method converges on a solution by computing Si+1 = Si * (P / R:SiP). For the novel Corona Method, we compute the corona using iterative backprojection reconstruction using ratios on only the target volumes T, T’ = R:S0T, and then superimpose the organ at risk, P’ = T’ + OAR. Iterative backprojection on P’ is then performed by successive addition correction, Si+1 = Si‐1 + (SO − Si), where SO is the sinogram of P’. Any negative intensities are set to zero during the iteration process. Target coverage is improved using the Compensated Phantom Method by computing a revised target based on the results Z of the process described thus far. We then define a new target T” = T’/Z and superimpose the OAR, such that P” = T” + OAR. The final procedure is to perform additive iteration constrained to contain only positive intensities. Results: Ratio Method: Target and boost areas receive their respective dose goals but the organ at risk unacceptable. Corona Method : OAR acceptable but poor target coverage. Compensated Phantom Method : Acceptable target coverage and the OAR goal is achieved. Conclusion: Dose image backprojection with initial corona calculation significantly reduces the dose to the organ at risk while maintaining acceptable target coverage.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here