SU‐GG‐T‐608: Modeling Scatter‐To‐Primary Dose Ratio for Megavoltage Photons
Author(s) -
Wang K,
Zhu T
Publication year - 2010
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3469009
Subject(s) - monte carlo method , photon , data set , dosimetry , monitor unit , beam (structure) , physics , optics , nuclear medicine , computer science , mathematics , statistics , medicine
Purpose : In radiation oncology, measured x‐ray dose ( TPR or PDD ) at central axis are usually tabulated and used for clinical dose calculation. An efficient technique to examine these datasets whether contain systematic or random errors can be useful to maintain the accuracy of the data and improve patient treatment. An optimization routine employing a newly developed scatter‐to‐primary dose ratio ( SPR ) can be used to examine the accuracy of clinical beam data. The SPR formula proposed in this study is more accurate than that published previously, especially for low energy beams, because the effect of backscattered photon is taken into account. Method and Materials : A formula, involving three fitting parameters, is formulated to express the SPR . Because of absence of measurement error and the ability separating scattering and primary dose component, Monte Carlo (MC) data is used to extract these parameters. Based on the results of fitting MC data, we are able to obtain a set of empirical expression for the fitting parameters as a function of photon beam attenuation coefficients. These empirical expressions are implemented into our optimization routine, used to match measured beam data. Results : In terms of fitting accuracy for MC, the new SPR formula outperforms the previously published form. For clinical data study, our optimization routine successfully fits the measured photon beam data with depth and field size up to 40 cm for a large variety of linear accelerators (6 to 25 MV) and Co60 machine from different manufactures. The maximum error among these fits is 2.8 % for Co60 and better than 2 % for all other energies. Conclusion : Because of the high fitting accuracy, our optimization routine, incorporating the new SPR form, can serve as a tool performing quality control of x‐ray beam data in a fast‐pace clinical environment.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here