SU‐GG‐T‐201: Managing the Transition of IMRT QA Systems: From Radiographic Films to Detectors Array
Author(s) -
NevesJunior W,
Alves T,
Casagrande T,
Mancini A,
Pelosi E,
Haddad C
Publication year - 2010
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3468590
Subject(s) - quality assurance , imaging phantom , ionization chamber , dosimetry , nuclear medicine , linear particle accelerator , radiation treatment planning , radiography , coronal plane , medical physics , materials science , mathematics , medicine , radiation therapy , physics , optics , ionization , beam (structure) , radiology , ion , external quality assessment , pathology , quantum mechanics
Purpose : This study aims to evaluate the results of intensity modulated radiotherapy quality assurance (IMRT QA) obtained with the MatriXX Evolution system and compare them to those previously obtained with, in order to implement this new technology in clinical routine, turning it more practical and still reliable. Materials and methods : Three moments are being compared: (i) an “Old Standard”, when the IMRT QA was performed with a polymethylmethacrylate slabs phantom, ionization chamber and radiographic film in the coronal plane; (ii) “Transition”, when the use of MatriXX system was initiated, but still in conjunction with a film simultaneously and (iii) “New Standard”, with MatriXX alone. The measurements were performed in linear accelerators with step‐and‐shoot technique. OmniPro‐I'mRT software was used to obtain the measured dose distributions (MatriXX and digitized film) and to compare them to calculated dose distributions ‐ by two treatment planning systems, Helax TMS and Oncentra MasterPlan. The percentage of points that meet the criteria of 3% dose difference and 3mm distance to agreement using the gamma function (γ) was evaluated, denoted here as γ≤1. Results : The IMRT QA with the “Old Standard” (141 patients) provides 2.1%±1.6% average dose difference and 95%±5% of γ≤1. During the “Transition” (48 patients), the average dose difference was 2.6%±1.4%, and the γ≤1 were 98%±2% with MatriXX and 99%±1% with films. Finally, with the “New Standard” (116 patients) a 2.2%±1.8% average dose difference and 98%±2% of γ≤1 where obtained. Conclusions : The results indicate a learning curve along this whole experience, as well as equivalence between film and MatriXX measured dose distributions, despite the distinct resolutions. IMRT QA performed with MatriXX Evolution system can properly substitute the ionization chamber and film pattern, reducing the time spent to execute these processes in clinical routine.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here