SU‐FF‐T‐386: Validation of the Delta4 Dosimetry Phantom Against Ionometric Measurements
Author(s) -
Calvo O,
Gutiérrez A,
Stathakis S,
Mavroidis P,
Moral S,
Esquivel C,
Shi C,
Papanikolaou N
Publication year - 2009
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3181867
Subject(s) - tomotherapy , ionization chamber , nuclear medicine , imaging phantom , dosimetry , pinnacle , radiation treatment planning , quality assurance , dose profile , medicine , medical physics , radiation therapy , physics , surgery , ionization , ion , external quality assessment , pathology , quantum mechanics
Purpose: To validate a 3D dose calculation methodology used by the 3D Delta4™ phantom. Method and materials: Measurements were performed using a TomoTherapy™ HiArt™ and a Varian 2300C/D™. A pinpoint PTW N31006 chamber with a sensitive volume of 0.016cc was use for point dose measurements. The Delta4™ phantom (ScandiDos AB, Uppsala, Sweden) was modified to accept the ion chamber. Eight (n=8) quality assurance plans were created in both planning systems. Plans covered a range of doses from 20% to 160% (20% steps) of the prescribed dose using Pinnacle 3 TPS. Tomotherapy plans covered doses from 20% to 100% (20% steps) and a DQA plan with 200% of the prescribed dose. Treatment plan, DQA plan, DQA dose and structures were exported via DICOM RT to the Delta4™ software. At each delivery, a point dose measurement was recorded. Point dose measurements and calculated doses by the Delta4™ were compared against the calculated dose of the treatment planning systems. Furthermore, point dose measurements were compared against the Delta4™ calculated dose. Results: The preliminary results showed that chamber measurements agree with Pinnacle 3 within 0.3% to 2%. The Delta4™ system have an excellent response at higher doses while at lower doses the percent differences are compared to the ones obtained by the pinpoint chamber. Tomotherapy delivery results showed good agreement between the pinpoint chamber and Delta4™ with percent differences ranging from 0.03% to 5% with better response at high doses Conclusion: Delta4™ showed good agreement with the pinpoint chamber in the dose calculation, which shows that the system 3D dose calculation methodology is able to predict the same or better doses compared to the pinpoint chamber measurements.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here