SU‐FF‐T‐314: Patient‐Specific Quality Assurance Techniques for RapidArc Radiotherapy
Author(s) -
O'Daniel J,
McMahon R,
Chang Z,
Yan H,
Sakhalkar H,
Das S,
Wu Q,
Yin F
Publication year - 2009
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3181794
Subject(s) - ionization chamber , quality assurance , nuclear medicine , dosimetry , materials science , radiation therapy , sagittal plane , dose rate , medical physics , medicine , ion , physics , radiology , external quality assessment , pathology , quantum mechanics , ionization
Purpose: To develop a patient‐specific quality assurance protocol for RapidArc™ radiotherapy. Method and Materials: Following the commissioning of RapidArc™ treatment delivery, we tested several methods of patient‐specific quality assurance (QA). Ion chamber and film measurements were used as the “gold standard”. One ion chamber measurement was taken in a region of low‐dose gradient. Four film measurements were taken in three planes: axial, sagittal, coronal, and a coronal180° (testing setup uncertainty). Measurements were also taken using the Matrixx® device (IBA Dosimetry, Bartlett, TN) in two orientations, sagittal and coronal. The Matrixx® consists of a 24×24cm 2 grid of ion chambers, providing a 2D absolute dose comparison. All plans were designed and calculated using Eclipse (Varian Medical Systems, Palo Alto, CA). Eleven patient plans were evaluated, consisting of 4 brain, 1 head&neck boost, and 6 prostate radiotherapy treatments. Results: Ion chamber measurements were typically ∼2% greater than those predicted by Eclipse. Using a gamma index with 3%, 3mm, 5% threshold criteria, 88% of films had gamma pass rates of >90%. Three of 7 axial films failed, but passed with thresholds of 30–40%. Similarly, Matrixx® measurements had gamma pass rates of >90% for 100% of measurements. The average reading of 4 Matrixx ion chambers was significantly correlated with the single ion chamber results (Pearson correlation coefficient = 0.839, p = 0.002), and the 2D Matrixx results agreed well with film measurements (gamma pass rate >90% for 10 of 11 comparisons). Conclusion: These QA measurements produced good agreement between planned and measured doses. Through utilizing ion chamber and film measurements, we have gained confidence in the Matrixx® for RapidArc QA. The chosen passing criteria were +/−3% for ion chamber readings, and gamma passing rates of >90% for film and Matrixx® with 3%, 3mm, and 5% threshold.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom