SU‐FF‐I‐08: Ramping and Overscanning Contributions to Patient Dose in Helical CT
Author(s) -
Tien C,
Winslow J,
Hintenlang D
Publication year - 2009
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3181127
Subject(s) - collimated light , scanner , dosimeter , materials science , nuclear medicine , dosimetry , rotation (mathematics) , optics , beam (structure) , multislice , physics , medicine , mathematics , laser , geometry
Purpose: In helical computed tomography (CT) scans, slices adjacent to the clinical volume of interest, known as z overscanning lengths, are acquired for reconstruction. Additionally, CT machines undergo a transient power state, known as tube ramping, to reach full power. This investigation will characterize the dependence of tube ramping time and z overscanning length on pitch, rotation time, beam collimation, and reconstruction slice width. These relationships will be utilized to determine the contributions to patient dose. Methods and Materials: A real‐time fiber optic coupled dosimetry system with 10 ms temporal resolution is interfaced with three organic scintillators with 0.157 mm 3 volumes. A Siemens Somatom Sensation 16 multislice helical CT scanner is used for the study with pitches varying from 0.5 to 1.5; rotation speeds between 0.5 and 1.5 seconds per rotation; reconstructed slice widths of 2.0 to 10.0 mm; and beam collimation of either 12 or 24 mm. One dosimeter remains stationary while two dosimeters are translated through the beam to provide real‐time profiles of the x‐ray field. Results: The ramping time was found to be independent of pitch and represented 0.5% of the total scan time. The z overscanning length was found to be monotonically increasing with pitch and represents anywhere from 14.3% to 35.9% of the total scan length. Conclusions: While ramping time remained constant, it has been demonstrated that z overscanning length is heavily dependent upon pitch — a high pitch added 7.1 cm to a 20.3 cm clinical scan length. Both z overscanning and ramping time must be taken into consideration when planning and managing patient dose. The effect of rotation time, beam collimation, and reconstruction slice width on ramping time and z overscanning length will be investigated in future work.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom