Absolute and relative dose measurements with Gafchromic™ EBT film for high energy electron beams with different doses per pulse
Author(s) -
Fiandra Christian,
Ragona Riccardo,
Ricardi Umberto,
Anglesio Silvia,
Giglioli Francesca Romana
Publication year - 2008
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.3005975
Subject(s) - dosimetry , dosimeter , ionization chamber , materials science , dose profile , pulse (music) , absorbed dose , electron , percentage depth dose curve , linear particle accelerator , ionization , optics , nuclear medicine , beam (structure) , detector , physics , nuclear physics , medicine , ion , quantum mechanics
The authors have evaluated the accuracy, in absolute and relative dose measurements, of the Gafchromic™ EBT film in pulsed high‐energy electron beams. Typically, the electron beams used in radiotherapy have a dose‐per‐pulse value of less than 0.1 mGy /pulse. However, very high dose‐per‐pulse electron beams are employed in certain linear accelerators dedicated to intraoperatory radiation therapy (IORT). In this study, the absorbed dose measurements with Gafchromic™ EBT in both low (less than 0.3 mGy per pulse) and high (30 and 70 mGy per pulse) dose‐per‐pulse electron beams were compared with ferrous sulfate chemical Fricke dosimetry (operated by the Italian Primary Standard Dosimetry Laboratory), a method independent of the dose per pulse. A summary of Gafchromic™ EBT in relative and absolute beam output determination is reported. This study demonstrates the independence of Gafchromic™ EBT absorption as a function of dose per pulse at different dose levels. A good agreement (within 3%) was found with Fricke dosimeters for plane‐base IORT applicators. Comparison with a diode detector is presented for relative dose measurements, showing acceptable agreement both in the steep dose falloff zone and in the homogeneous dose region. This work also provides experimental values for recombination correction factor ( K sat ) of a Roos (plane parallel) ionization chamber calculated on the basis of theoretical models for charge recombination.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here