SU‐GG‐T‐244: Commissioning of Flattening Filter Free Varian 600C Linear Accelerator
Author(s) -
Stathakis S,
Gutiérrez A,
RassiahSzegedi P,
Mavroidis P,
Papanikolaou N
Publication year - 2008
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.2961996
Subject(s) - flattening , linear particle accelerator , percentage depth dose curve , beam (structure) , monitor unit , nuclear medicine , dosimetry , ionization chamber , field size , optics , fraction (chemistry) , physics , chemistry , medicine , ion , quantum mechanics , astronomy , ionization , organic chemistry
Purpose: The purpose of this study is twofold: a)to determine the dosimetric features of an unflattened 6‐MV photon beam of a Varian 600C linac; and b)to compare the beam characteristics of the unflattened beam against the flattened one. Method and Materials: Comparison of measurements in water was performed for a Varian 600C linear accelerator with and without flattening filter. The measurements consisted of dose profiles (inplane and cross plane), percent depth doses, and photon beam output. The field sizes measured varied from 2cm×2cm to 40cm×40cm. All measurements were taken at 100cm SSD with an A1SL ion chamber. Results: Our results showed a significant (about twofold) increase in the dose rate for all field sizes. Also, a steeper reduction in depth doses was observed when the flattening filter was removed. No decrease was observed in the out‐of‐field dose, but there was an increase of the dose in the build‐up region that can be attributed to the increase of contaminant electrons. Conclusion: . Our study revealed that, for recent radiotherapy techniques, removing the flattening filter could be beneficial since there is a an increased dose rate and higher output therefore less monitor are required for each fraction leading to reduced whole body dose equivalent. On the other hand, higher skin doses are observed per monitor unit delivered but the overall skin dose for a complete treatment is comparable to the one with flattening filter since less monitor unit are required per fraction.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here