SU‐FF‐T‐450: Verification of the Source Tech Medical Model STM1251 I‐125 Air‐Kerma Strength in Japan
Author(s) -
Sumida I,
Takahashi Y,
Ito A,
Nose T,
Oguchi M,
Yamashita T,
Yoshioka Y,
Inoue T
Publication year - 2007
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.2761175
Subject(s) - kerma , nuclear medicine , mean value , standard deviation , range (aeronautics) , mathematics , medicine , materials science , statistics , composite material , dosimetry
Purpose: We investigated the difference between measured and manufacturer's stated source strength in a single model STM1251125 I seeds. Materials and methods: A well‐type ionization chamber with a single seed holder was used to measure source strength of 2412125 I seeds before implant for 34 patients. The air kerma strength was only used 0.279 U for all cases. The mean source strength used for each patient was compared with the manufacture stated value. The deviation from the measured value was compared with the tolerance level of the recommendation of AAPM TG‐40. Results: Measured source strength was higher than the manufacture stated value with median 1 % (range: −2 to 5 %). Sixteen seeds of all 2412 seeds (0.7 %) were over 5 % different from the manufacture stated value. Standard deviation from the mean value was median 2.2 % (range: 1.1 to 2.5 %) in each patient. Conclusion: A single seed assay was performed for the model STM1251 seed. The manufacture stated strength of the125 I seeds agreed well with the measured value. Our study was helpful as guidance for the permanent prostate implant using the model STM1251 seed. (Submission Withdrawn).